MS-PC 2015: Performance Specifications for Instrumentation Systems Designed to Measure Radon Gas in Air Page 16 of 20 Proportionality (Sec. 7.3) Requirements: Each IPE = 0 ± 25% and COV ≤ 15% for each set of five devices, and the difference between the average IPE(low concentration) and the average IPE(high concentration) must be between −15% and +15%. Note that a test in either a high radon concentration or a high integrated radon concentration is required (not both). Low Concentration Exposure: Temperature _______ range 18 – 24°C (65 – 75°F) Rn Conc __________ range 222 – 555 Bq/m 3 (6 – 15 pCi/L) RH______ range 10 – 55% Serial number Date-time start Date-time end Temp RH Device result STAR conc IPE COV High Concentration Exposure: Temperature _______ range 18 – 24°C (65 – 75°F) Rn Conc __________ range 1110 – 2220 Bq/m 3 (30 – 60 pCi/L) or at least 3 times the low concentration, whichever is greater RH______ range 10 – 55% Serial number Date-time start Date-time end Temp RH Device result STAR conc IPE COV